Jawa Barat - Bekasi

AOIP GREENLINE 6000 Hand-held industrial combustion gas analyzer with 6 sensors

Harga :
CALL
Merk :
AOIP
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Dilihat :
103 Kali
Update Terakhir :
22-03-2016 15:54

Dapatkan Spesifikasi Harga AOIP GREENLINE 6000 Hand-held industrial combustion gas analyzer with 6 sensors Produk Terbaik dari Andalan Prima General Supplier,Test & Measuring Instrument,Laboratory Equipment , Glassware & Chemical

Detail Produk

 
Measured parameter Range Resolution GreenLine 6000
O2 From 0 to 25% vol. 0.1 % EC
CO From 0 to 8000 ppm 1 ppm EC
CO From 0 to 20000 ppm 1 ppm EC
CO LOW From 0 to 500 ppm 0.1 ppm EC
CO From 0 to 10% 0.01% EC
CO2 From 0 to 99.9% 0.1% Calc
NO From 0 to 4000 ppm 1 ppm EC
NO LOW From 0 to 500 ppm 0.1 ppm EC
NOx From 0 to 4000 ppm 1 ppm Calc
NO2 From 0 to 1000 ppm 1 ppm EC
NO2 LOW From 0 to 100 ppm 0.1 ppm EC
SO2 From 0 to 4000 ppm 1 ppm EC
CxHy From 0 to 5% 0.01% Pellistore
H2S From 0 to 1000 ppm 1 ppm EC
Efficiency From 1 to 99.9% 0.1% Calc
Excess air From 1 to infinite 0.01% Calc
Delta P ±100 hPa 0.01 hPa
Tair/Tgas -10...100°C / 0...1000°C 0.1°C Pt100 / Tc K
Calc

Further features

Zero calibration Automatic calibration procedure at instrument power-on. Fresh air inlet with electrovalve and separate pneumatic circuit
Self-diagnosis Sensor efficiency test with diagnostic page
Gas level alarms Programmable from PC with GasConfig software
Sampling pump 2.2 l/min - 220 mbar with electronic flow controller
Smoke measurement Using the heated probe or the optional external manual pump
Index memory store and printout capability as standard

General specifications

Size MCU: 455 x 205 x 365 mm
RCU: 115 x 90 x 330 mm
Weight MCU: 10 kg
RCU: 0.9 kg
Power supply 110 / 230 VAC (50/60 Hz)
Battery Type: Rechargeable 7.2 Ah
Battery life: 10 hours continuous use
Communication ports RS 232
Bluetooth
Internal test memory Up to 1,000 complete analysis data points structured by Tags

Environmental specifications

Operating reference range From -5°C to +45°C (up to 50°C for short time)
Storage temperature limits From -20 to +60°C (3 months max. at temperatures exceeding the operational limits)

Permintaan Penawaran