3 channel inputs for the simultaneous recording of Leakage Current , Load Current and Voltage
• Power Quality analysis• Large capacity for storing 60,000 data points
• Lowpass Filter will filter out the harmonics
• LED flickers when the preset current/voltage value is exceeded
• CALL:Confirmation of recorded data
• Selection of One-time mode or Endless mode
• Non Volatile Memory
• Battery power indicator
• The userfriendly PC software "KEW LOG Soft 2" is supplied
• Continuous measuring time:Approx. 10 days (Alkaline Battery)

- 3 channel inputs for the simultaneous recording of Leakage Current, Load Current and Voltage
 - Large Capacity for storing 60,000 data points
    
- 60,000 data points can be recorded when 1ch is used, and when all the three channels are used, 20,000 data points per channel can be recorded
 
 - Lowpass Filter will filter out the harmonics (Cutoff Frequency = Approx. 160Hz)
 - LED flickers when the preset current / voltage value is exceeded
 - CALL: Confirmation of recorded data
    
- The following can be displayed: number of recorded data points, (max+ min+ peak) value for each channel complete with time/date information in the Normal recording mode. (Detected values (i.e. when values are outside preset limits ) can be displayed in other recording modes)
 - RECALL: The last 10 recorded data points including time/date can be recalled on the logger display
 
 - Selection of One-time mode or Endless mode
    
- One-time on: Recording will stop when memory is used up
 - One-time off: Overwrite the old data, and store the latest data
 
 - Non Volatile Memory
Recorded data will be retained even if the batteries are exhausted or replaced due to the presence of a nonvolatile memory (guaranteed for 10 years) - Battery power indicator
Indicates battery voltage in 4-levels
(It is possible to use the logger for a further approx. 24 hours even after the warming symbol is flashing) - The userfriendly PC software "KEW LOG Soft 2" is supplied
    
- Supplied with the user friendly software "KEW LOG Soft 2". This permits editing, analysis and graphical display of data.
 - The recorded data is downloadable onto a PC via USB cable.
 - Variation of the measured voltage and current data can be confirmed simultaneously on the PC display monitor. (only on KEW 5020)
 - Simplified Power Integration
(The "KEW LOG Soft 2" uses current and voltage recorded to calculate the integral power consumption) 
 - Continuous measuring time: Approx. 10 days (Alkaline Battery)
 
Normal Recording Mode
(AC 50/60Hz, Sine wave, Input: 10% or more of the range at CH1)
| Range | RMS Accuracy | 
|---|---|
| 100.0mA | ±2.0%rdg±0.9%f.s.+ Accuracy of Sensor | 
| Other ranges | ±1.5%rdg±0.7%f.s.+ Accuracy of Sensor | 
| Crest Factor | 2.5 or less: RMS accuracy (sine) + 2%rdg + 1%f.s. | 
*Max, Min and Instant Peak values in Normal Recording mode are just reference values:
their accuracies aren't guaranteed.
Trigger Recording Mode (AC 50/60Hz sine wave)
| Range | Accuracy | 
|---|---|
| 100.0mA | ±3.5%rdg±2.2%f.s.+ Accuracy of Sensor | 
| Other ranges | ±3.0%rdg±2.0%f.s.+ Accuracy of Sensor | 
Capture Recording Mode
| Range | Accuracy | 
|---|---|
| 100.0mA | ±3.0%rdg±1.7%f.s.+ Accuracy of Sensor | 
| Other ranges | ±2.5%rdg±1.5%f.s.+ Accuracy of Sensor | 
| Recording mode | Normal, Trigger, Capture, Power Quality Analysis | 
|---|---|
| Operating system | Successive Approximation (CH1 single synchronized sampling)  | 
        
| Rated max. working voltage | AC 9.9Vrms, 14V peak value | 
| Number of input channel | 3ch | 
| Measuring method | True RMS | 
| RMS measuring interval | approx. 100ms | 
| Sampling interval | Normal/Trigger mode : approx. 1.65ms/CH  Capture mode : approx. 0.55ms (waveform: at every 1.1ms) P.Q.A. mode : approx. 0.55ms  | 
        
| Low battery warning | Battery mark display (in 4 levels) | 
| Over-range indication | "OL" mark is displayed when exceeding the measuring range | 
| Auto power off | Power-off function operates automatically after a switch remains for 3min. (when recording is stopped) | 
| Possible measurement time | Approx. 10 days (with alkaline LR6 batteries) | 
| Applicable standards | IEC 61010-1:2001 CAT III 300V Pollution degree2 IEC 61326 (EMC standard)  | 
        
| Dimensions | 111(L)×60(W)×42(D)mm | 
| Weight | Approx. 265g | 
| Included Accessories | Alkaline battery LR6×4pcs, 9118 (Carrying case) PC software "KEW LOG Soft 2", 7148 (USB cable) Instruction manual, Quick manual, Install manual, USB Notice sheet  | 
        
| Optional Accessories | 8146/8147/8148 (Leakage & Load current Clamp Sensor), 8121/8122/8123 (Load current Clamp Sensor), 8129/8130 (Flexible Clamp Sensor), 8309 (Voltage Sensor: only KEW 5020), 8320 (AC Adapter), 9135 (Carrying bag), 9119 (Hard carry case), 7185 (Extension cable for sensor)  | 
        
                        