Memmert CO2 Incubator ICO50
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Update Terakhir :
03-06-2016 09:16
Dapatkan Spesifikasi Harga
Memmert CO2 Incubator ICO50 Produk Terbaik dari
Andalan Prima General Supplier,Test & Measuring Instrument,Laboratory Equipment , Glassware & Chemical
Detail Produk
Properties
Control of standard components
| ControlCOCKPIT |
adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
| Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
| Humidity control |
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 % |
| Setting range active humidity control |
40 to 97 % rh and rh-Off |
| Setting accuracy humidity |
0.5 % rh |
| CO2 control |
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation |
| Adjustment range CO2 |
0 to 20 % CO2 |
| Variation in time CO2 |
+/- 0.2 % CO2 |
| Adjustment range O2 |
1 to 20 % O2 |
| Setting accuracy O2 |
0.1 % O2 |
Temperature
| Setting accuracy temperature |
0.1 °C |
| Setting temperature range |
+18 to +50 °C |
| Working-temperature range |
5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors) |
Control technology
| Function SetpointWAIT |
the process time does not start until the set temperature is reached |
| Language setting |
German/English/Spanish/French |
Communication
| Documentation |
programme stored in case of power failure |
| Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
| Alarm |
visual and acoustic |
Heating concept
| large-area multi-function heating system on four sides with additional door and back heating to avoid condensation |
Standard equipment
| Scope of delivery |
incl. works calibration certificate for +37°C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); Standard equipment for units with O2 control |
| Scope of delivery |
incl. works calibration certificate for +37°C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control |
| Scope of delivery |
incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units |
| Internals |
1 pc stainless steel shelf, perforated |
| Internals |
1 pc stainless steel water dish |
| Door |
fully insulated stainless steel door with 2-point locking (compression door lock) |
| Door |
inner glass door with opening (8 mm Ø) to take gas sample |
| Housing |
rear zinc-plated steel |
| Interior |
material 1.4301 (ASTM 304), corrosion resistant |
Stainless steel interior
| Max. loading of chamber: |
75 kg |
| Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 400 x 425 x 330 mm |
| Volume |
56 l |
Electrical data
| Voltage Electrical load |
230 V, 50/60 Hz
approx. 1000 W |
| Voltage Electrical load |
115 V, 50/60 Hz
approx. 1000 W |
Packing/shipping data
| the appliances must be transported upright |
| Customs tariff number |
8419 8998 |
| Country of origin |
Federal Republic of Germany |
| WEEE-Reg.-No. |
DE 66812464 |
| Dimensions approx incl. carton |
B x H x T: 730 x 950 x 640 mm |
| Net weight |
approx. 55 kg |
| Gross weight carton |
approx. 74 kg |