MEMMERT Universal Oven UN55
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Update Terakhir :
01-06-2016 09:26
Detail Produk
Properties
Control of standard components
ControlCOCKPIT |
adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Temperature
Set temperature range in °C |
min.10 above ambient up to +300 |
resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED) |
Control technology
Function HeatBALANCE |
adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Language setting |
German/English/Spanish/French |
Calibration |
three freely selectable temperature values |
adjustable parameters |
temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime |
Ventilation
Fan |
forced air circulation by 2 quiet air turbines, adjustable in 10 % steps for each segment individually |
Fresh air admixture |
adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually |
Vent |
vent connection with restrictor flap |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
Temperature control |
overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY |
additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system |
for fault analysis |
Alarm |
visual and acoustic |
Standard equipment
Scope of delivery |
incl. works calibration certificate for +160°C |
Door |
fully insulated stainless steel doors with 2-point locking (compression door lock) |
Housing |
rear zinc-plated steel |
Interior |
easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Internals |
2 stainless steel grids |
Installation |
on lockable castors |
Stainless steel interior
Max. loading of chamber: |
300 kg |
Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 1040 x 1200 x 600 mm |
Volume |
749 l |
Textured stainless steel casing
w(D) x h(E) x d(F): 1224 x 1726 x 784 mm |
Electrical data
400 V, 3 phases, 50 Hz / approx. 7000 W |
Packing/shipping data
the appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
B x H x T: 1330 x 1910 x 1050 mm |
Net weight |
approx. 217 kg |
Gross weight carton |
approx. 288 kg |