MEMMERT CO2 Incubator ICO105
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Update Terakhir :
03-06-2016 09:19
Detail Produk
Properties
Control of standard components
ControlCOCKPIT |
adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Humidity control |
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 % |
Setting range active humidity control |
40 to 97 % rh and rh-Off |
Setting accuracy humidity |
0.5 % rh |
CO2 control |
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation |
Adjustment range CO2 |
0 to 20 % CO2 |
Variation in time CO2 |
+/- 0.2 % CO2 |
Adjustment range O2 |
1 to 20 % O2 |
Setting accuracy O2 |
0.1 % O2 |
Temperature
Setting accuracy temperature |
0.1 °C |
Setting temperature range |
+18 to +50 °C |
Working-temperature range |
5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors) |
Control technology
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Language setting |
German/English/Spanish/French |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Alarm |
visual and acoustic |
Heating concept
large-area multi-function heating system on four sides with additional door and back heating to avoid condensation |
Standard equipment
Scope of delivery |
incl. works calibration certificate for +37°C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); Standard equipment for units with O2 control |
Scope of delivery |
incl. works calibration certificate for +37°C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control |
Scope of delivery |
incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units |
Internals |
2 pcs stainless steel shelves, perforated |
Internals |
1 pc stainless steel water dish |
Door |
fully insulated stainless steel door with 2-point locking (compression door lock) |
Door |
inner glass door with opening (8 mm Ø) to take gas sample |
Housing |
rear zinc-plated steel |
Interior |
material 1.4301 (ASTM 304), corrosion resistant |
Stainless steel interior
Max. loading of chamber: |
90 kg |
Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 560 x 480 x 400 mm |
Volume |
107 l |
Electrical data
Voltage Electrical load |
230 V, 50/60 Hz
approx. 1500 W |
Voltage Electrical load |
115 V, 50/60 Hz
approx. 1500 W |
Packing/shipping data
the appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
B x H x T: 800 x 1030 x 800 mm |
Net weight |
approx. 75 kg |
Gross weight carton |
approx. 100 kg |