MEMMERT Incubator IF750plus
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Update Terakhir :
03-06-2016 10:55
Detail Produk
Properties
Control of standard components
ControlCOCKPIT |
adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Temperature
min. 10°C above ambient up to +80°C |
resolution of display for setpoint and actual temperature values |
0.1°C |
Control technology
Function HeatBALANCE |
adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Language setting |
German/English/Spanish/French |
Calibration |
three freely selectable temperature values |
Sterilisation |
fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load |
adjustable parameters |
temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime |
Ventilation
Fan |
forced air circulation by 2 quiet air turbines, adjustable in 10 % steps for each segment individually |
Fresh air admixture |
adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually |
Vent |
vent connection with restrictor flap |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
Temperature control |
overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY |
additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system |
for fault analysis |
Alarm |
visual and acoustic |