Control of standard components
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Humidity |
integrated humidity limit control (88-97 % rh) with digital display of relative humidity - resolution of display 0.5 %, setting accuracy 1 % |
CO2 |
CO2-supply via sterile filter |
CO2 |
intrinsic adjustment of altitude above sea level |
CO2 |
CO2-supply interruption upon door opening |
CO2 |
digital electronic CO2-control with automatic zeroing (infrared measuring system) - adjustment range 0 - 10 % - resolution 0.1 % |
Controller |
digital display of all set parameters, such a temperature, weekdays, time, CO2, humidity and set-up values - language to be chosen via set-up |
Controller |
Electronic microprocessor temperature controller with auto-diagnostic system |
Controller |
digital display of setpoit and actual values (LED) for temperature (resolution 0.1°C) and CO2 concentration (0.1 %) |
Timer |
digital 7-day programme timer with real time clock, precise minute setting |
Temperature
min. 8°C above ambient up to +50°C |
Control technology
Calibration |
three freely selectable temperature values, 2-point-calibration for humidity: 20 and 90 %, 3-point calibration for CO2: 5 %, 7 %, 10 % |
Ventilation
uniform atmosphere and temperature distribution owing to enclosed non-turbulent ventilation system in working chamber |
Communication
Documentation |
programme stored in case of power failure |
Safety
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
Temperature control |
overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY |
additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system |
integral fault diagnostics for temperature, CO2 and humidity limit control |
Alarm |
visual and acoustic |
Alarm |
Acoustic alarm: Over- and undertemperature, over-CO2 and empty gas cylinder, open door, underhumidity and empty water tank (with optional humidity module) |