MEMMERT CO2 Incubator (med) INCO153med
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Update Terakhir :
06-06-2016 09:13
Detail Produk
Properties
Control of standard components
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Humidity |
integrated humidity limit control (88-97 % rh) with digital display of relative humidity - resolution of display 0.5 %, setting accuracy 1 % |
CO2 |
CO2-supply via sterile filter |
CO2 |
intrinsic adjustment of altitude above sea level |
CO2 |
CO2-supply interruption upon door opening |
CO2 |
digital electronic CO2-control with automatic zeroing (infrared measuring system) - adjustment range 0 - 10 % - resolution 0.1 % |
Controller |
digital display of all set parameters, such a temperature, weekdays, time, CO2, humidity and set-up values - language to be chosen via set-up |
Controller |
Electronic microprocessor temperature controller with auto-diagnostic system |
Controller |
digital display of setpoit and actual values (LED) for temperature (resolution 0.1°C) and CO2 concentration (0.1 %) |
Timer |
digital 7-day programme timer with real time clock, precise minute setting |
Temperature
min. 8°C above ambient up to +50°C |
Control technology
Calibration |
three freely selectable temperature values, 2-point-calibration for humidity: 20 and 90 %, 3-point calibration for CO2: 5 %, 7 %, 10 % |
Ventilation
uniform atmosphere and temperature distribution owing to enclosed non-turbulent ventilation system in working chamber |
Communication
Documentation |
programme stored in case of power failure |
Safety
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
Temperature control |
overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY |
additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system |
integral fault diagnostics for temperature, CO2 and humidity limit control |
Alarm |
visual and acoustic |
Alarm |
Acoustic alarm: Over- and undertemperature, over-CO2 and empty gas cylinder, open door, underhumidity and empty water tank (with optional humidity module) |
Heating concept
large-area multi-function heating system on four sides with additional door and back heating to avoid condensation |
Standard equipment
Scope of delivery |
incl. works calibration certificate for +37°C |
Scope of delivery |
chip-card (STERICard) for sterilisation of working chamber with fixed values (4 hours/160°C), without removal of sensors and mountings |
Door |
fully insulated stainless steel door with 2-point locking (compression door lock), lockable |
Door |
inner glass door with opening (8 mm Ø) to take gas sample |
Housing |
rear zinc-plated steel |
Interior |
easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing, material 1.4301 (ASTM 304), hermetically welded |
Internals |
3 perforated stainless steel shelves and 1 stainless steel water tray |
Stainless steel interior
Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 480 x 640 x 500 mm |
Volume |
153 l |
Textured stainless steel casing
w(D) x h(E) x d(F): 630 x 938 x 650 mm |
Electrical data
Voltage Electrical load |
230 V, 50/60 Hz
approx. 1500 W |
Packing/shipping data
the appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
B x H x T: 830 x 1300 x 800 mm |
Net weight |
approx. 80 kg |
Gross weight carton |
approx. 106 kg |