MEMMERT Climate chamber ICH750L
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Update Terakhir :
06-06-2016 09:53
Detail Produk
Properties
Control of standard components
ControlCOCKPIT |
adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Humidity |
humidity supply with distilled water from external tank by self-priming pump |
Humidity |
humidity supply with distilled water from external tank by self-priming pump |
Humidity |
humidification by hot steam generator |
Humidity |
active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 % |
Light |
daylight and UV-light can be programmed individually via controller |
Light |
cold white light +/- 6 %, for UV-light +/- 10 % |
Light |
illumination unit consisting of 4 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm |
Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Temperature
from +10°C up to +60°C with humidity |
resolution of display for setpoint and actual temperature values |
0.1°C |
Control technology
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Language setting |
German/English/Spanish/French |
Calibration |
three freely selectable values each, temperature/humidity, for two freely selectable values CO2 |
adjustable parameters |
temperature (Celsius or Fahrenheit), fan speed, light, programme time, time zones, summertime/wintertime |
Ventilation
Fan |
forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
Temperature control |
over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY |
additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature |
Autodiagnostic system |
integral fault diagnostics for temperature and humidity control |
Alarm |
visual and acoustic |
Heating concept
no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior |
energy-saving use of CFC-free cooling/heating system (refrigerant R134a) |
highly efficient automatic defrosting system |
air jacket heating system for gentle all-around heating |
Standard equipment
Door |
inner glass doors |
Door |
fully insulated stainless steel doors with 2-point locking (compression door lock) |
Housing |
rear zinc-plated steel |
Internals |
2 stainless steel grids |
Installation |
on lockable castors |
Stainless steel interior
Max. loading of chamber: |
200 kg |
Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 1040 x 1200 x 600 mm |
Volume |
749 l |
Textured stainless steel casing
w(D) x h(E) x d(F): 1224 x 1950 x 834 mm |
Electrical data
Voltage Electrical load |
230 V, 50/60 Hz
approx. 1350 W |
Voltage Electrical load |
115 V, 50/60 Hz
approx. 1350 W |
Packing/shipping data
the appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
B x H x T: 1330 x 2150 x 1050 mm |
Net weight |
approx. 249 kg |
Gross weight carton |
approx. 319 kg |