Range | 0.00 to 9.99 ppt (g/L) | |
Resolution | 0.01 ppt | |
Accuracy | ±2% F.S. | |
Calibration | manual, 1 point through trimmer | |
Calibration Solution | HI 70038P | |
Temperature Compensation | automatic, 5 to 50°C (41 to 122°F) with ß=2%/°C | |
Setpoint | 0.70 to 3.50 ppt | |
Alarm | red LED (blinks when measurement differs from setpoint by: ±0.2 ppt |
|
Probe | HI 7632/2 with 2 m (6.6') cable and ½" thread (included) |
|
Power Supply | 12 Vdc power adapter (included) | |
Environment | 0 to 50°C (32 to 122°F); RH max 100% | |
Dimensions | 86 x 110 x 43 mm (3.4 x 4.3 x 1.7") | |
Weight | 215 g (7.6 oz.) |