DISPLAY | |
8 digits with Hz, kHz, MHz, GHz, s, ms, μs, and overflow | |
GATE TIME | |
Variable from 10ms to 10s, or 1 period of input depending on whichever is greater | |
ACCURACY | |
±(resolution ± timebase error) | |
CHANNEL A | |
Range | DC coupling: 0.01Hz ~ 120MHz AC coupling: 30Hz ~ 120MHz |
Sensitivity | 10mV typical, 50mV maximum |
Coupling | AC or DC, selectable |
Filter | Low pass, selectable in or out for channel A (-3dB point, nominal 100kHz) |
Impedance | 1MΩ // 40pF |
Attenuator | 1/1 or 1/20dB |
Trigger Level | -2.5VDC ~ +2.5VDC |
Adjustment Resolution | For frequency measurement, the maximum resolution is 100nHz for 1Hz and 0.1Hz for 100MHz inputs respectively. For period measurement, the maximum resolution is 10ns for 1Hz input, and 0.1x10-15s for 100MHz inputs respectively. |
Display Accuracy | 7 digits for 1 sec gate time 6 digits for 100ms gate time 5 digits for 10ms gate time |
Period Range | 8ns ~ 100ns (at least 7 digits displayed per gate time) |
CHANNEL B | |
Range | 50MHz ~ 2.7GHz for GFC-8270H 50MHz ~ 1.3GHz for GFC-8131H |
Sensitivity | <50mVrms (10mVrms typical) for GFC-8270H <40mVrms (10mVrms typical) for GFC-8131H |
Coupling | AC only |
TIME BASE | |
Aging Rate | 1ppm per month |
Temperature | 5ppm, 23°C + 5°C |
Line Variation | 0.005ppm for + 10% variation |
POWER SOURCE | |
AC 100V/120V/220V/230V ±10%, 50/60Hz | |
DIMENSIONS & WEIGHT | |
230(W) × 95(H) × 280(D) mm Approx. 2.2kg |
Features