Physical | |
---|---|
Dimensions (hxwxd) | Ø47mm, 203 mm long |
Weight | 0.8 kg |
Protection Class | IP68 |
Enclosure Material | Body - ABS, Titanium Temperature Probe with Polyurethane coated cable Cap - ABS, Polycarbonate/Polymethyl/methylacrylate |
Cable Entry | Integral Cable Gland |
Cable Length | 10 metres standard, 100 metres maximum |
Interface to Monitor | |
Power Supply | 12 VDC from Monitor |
Sensor Output | ModTechw², RS485 |
Environmental Data | |
Operating/Storage Temperature | 0 to 50ºC |
Location | Indoor/Outdoor |
Pressure Rating | 210 mWC |
Flow | No flow necessary |
Maintenance | |
Service | The optical cap for the sensor should be replaced every 24 months Periodic cleaning is required at a frequncy determined by the application. Our self cleaning shaft option is highly recommended, this eliminates manual cleaning between calibrations. |
Calibration | The sensor is inherently stable, calibration should be carried out to suit the application. Our recommendation is yearly |