Jawa Barat - Bekasi

HACH TU5400 sc Ultra-High Precision Low Range Laser Turbidimeter

Harga :
CALL
Merk :
HACH
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Dilihat :
67 Kali
Update Terakhir :
02-09-2016 14:15

Detail Produk

 

Specifications

Accuracy:±2% of reading plus 0.01 NTU from 0 to 40 NTU
Cable Length:1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration Method:For Formazin and Stablcal: 
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range

Custom calibration for up to 6-point calibrations
Certifications:CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian RCM Marking
Communication:System Check and RFID
Dimensions (H x W x D):249 mm x 268 mm x 190 mm
Fitting type:Sample quick connector: ¼-in. for ¼-in. tubing
Light Source:Class 2 laser product, with embedded 650 nm (EPA) or 850 nm (ISO), max. 1.0 mW Class 2 laser source (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Lower Limit of Detection (LOD):0.0001 NTU
Material:ASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40,
Thermoplastic Elastomer TPS-SEBS
Mounting Configurations:Wall Mount
Operating Humidity:Relative humidity: 5 to 95% at different temperatures, non-condensing
Operating Temperature Range:0 to 50 °C (32 to 122 °F)
Options:Process Head with Flow Sensor and Automatic Cleaning Module
Range:EPA:
0 to 700 NTU / FNU / TE/F / FTU
0 to 175 EBC

ISO:
0 to 1000 NTU / FNU / TE/F / FTU
0 to 250 EBC
Regulatory:EPA
Repeatability:Better than 1% of reading or ±0.0006 NTU on Formazin at 25 °C (77 °F), whichever is greater
Resolution:0.0001 NTU / FNU / TE/F / FTU / EBC
Response Time:T90<30 seconds at 100 mL/min
Sample Flow Rate:100 to 1000 mL/min; optimal flow rate: 200 to 500 mL/min
Sample Pressure:6 bar (87 psi) maximum, compared to air at sample temperature range from 2 to 40 °C (35.6 to 104 °F)
Sample Temperature:2 to 70 °C (35.6 to 158 °F)
Signal Average Time:5 to 90 seconds (default: 30 seconds)
Storage Conditions:-40 to 60 °C (-40 to 140 °F)
Stray Light:<10 mNTU
Units:NTU, FNU, TE/F, FTU, EBC
Verification:Liquid: Stablcal®, Formazin (0.1 to 40 NTU)
Dry: Glass Rod at<0.1 NTU
Warranty:1 year
Weight:5.95 lbs. (2.7 kg) (11lbs. (5.0 kg) with all accessories)

Permintaan Penawaran