Jawa Barat - Bekasi

Fluke 1730/US Three-Phase Energy Logger

Harga :
CALL
Merk :
FLUKE
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Barat - Bekasi
Dilihat :
67 Kali
Update Terakhir :
13-09-2016 10:49

Detail Produk

 
Accuracy Specifications
Parameter Range Resolution Intrinsic Accuracy at Reference Conditions (% of Reading + % of Full Scale)
Voltage 1000 V 0.1 V ±(0.2% + 0.01%)
Current iFlex1500-12 150 A 0.1 A ±(1% + 0.02%)
1500 A 1 A
iFlex3000-24 300 A 1 A ±(1% + 0.02%)
3000 A 10 A
iFlex6000-36 600 A 1 A ±(1.5% + 0.03%)
6000 A 10 A
i40s-EL clamp 4 A 1 mA ±(0.7% + 0.02%)
40 A 10 mA
Frequency 42.5 Hz to 69 Hz 0.01 Hz ±(0.1%)
Auxillary input ±10 VDC 0.1 mV ±(0.2% + 0.02%)
Voltage min/max 1000 V 0.1 V ±(1% + 0.1%)
Current min/max Defined by accessory Defined by accessory ±(5% + 0.2%)
Cosφ/DPF 0 ≤ Cosφ ≤ 1 0.01 ±0.025
Power factor 0 ≤ PF ≤ 1 0.01 ±0.025
THD on voltage 1000% 0.1% ±(2.5% ± 0.05)
THD on current 1000% 0.1% ±(2.5% ± 0.05)
Intrinsic Uncertainty ±(% of reading + % of range)1
Parameter Influence quantity iFlex1500-12 iFlex3000-24 iFlex6000-36 i40S-EL
150 A / 1500 A 300 A / 3000 A 600 A / 6000 A 4 A / 40 A
Active power P PF ≥ 0.99 1.2% + 0.005% 1.2% + 0.0075% 1.7% + 0.0075% 1.2% + 0.005%
0.5 < PF < 0.99 1.2% + 7 x (1-PF) +0.005% 1.2% + 7 x (1-PF) + 0.0075% 1.7% + 7 x (1-PF) + 0.0075% 1.2% + 10 x (1-PF) + 0.005%
Apparent power S, S fund. 0 ≤ PF ≤ 1 1.2% + 0.005% 1.2% + 0.0075% 1.7% + 0.0075% 1.2% + 0.005%
Reactive power 
N, Q fund.
0 ≤ PF ≤ 1 2.5% of measured apparent power
Additional uncertainty in % of range1 U > 250 V 0.015% 0.0225% 0.0225% 0.015%
1. Range = 1000 V x Irange
Reference conditions: 
Environmental: 23°C ± 5°C, instrument operating for at least 30 minutes, no external electrical/magnetic field, RH < 65% 
Input conditions: Cosφ/PF = 1, Sinusoidal signal f = 50 Hz / 60 Hz, power supply 120 V / 230 V ±10%
Current and power specifications: Input voltage 1 ph: 120 V / 230 V or 3 ph wye/delta: 230 V / 400 V 
Input current: I > 10% of Irange 
Primary conductor of clamps or Rogowski coil in center position 
Temperature coefficient: Add 0.1 x specified accuracy for each degree C above 28°C or below 18°C
Electrical Specifications
Power Supply
Voltage range 100 V to 500 V using safety plug input when powering from the measurement circuit
100 V to 240 V using standard power cord (IEC 60320 C7)
Power consumption Maximum 50 VA (max. 15 VA when powered using IEC 60320 input)
Efficiency ≥ 68.2% (in accordance with energy efficiency regulations)
Maximum no-load consumption < 0.3 W only when powered using IEC 60320 input
Mains power frequency 50/60 Hz ± 15%
Battery Li-ion 3.7 V, 9.25 Wh, customer-replaceable
On-battery runtime Four hours in standard operating mode, up to 5.5 hours in power saving mode
Charging time < 6 hours
Data Acquisition
Resolution 16-bit synchronous sampling
Sampling frequency 5120 Hz
Input signal frequency 50/60 Hz (42.5 to 69 Hz)
Circuit types 1-φ, 1-φ IT, Split phase, 3-φ delta, 3-φ wye, 3-φ wye IT, 3-φ wye balanced, 3-φ Aron/Blondel (2-element delta), 3-φ delta open leg, Currents only (load studies)
THD THD for voltage and current is calculated using 25 harmonics
Averaging period User selectable 1 s, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min
Demand interval User selectable 5 min, 10 min, 15 min, 20 min, 30 min
Data storage Internal flash memory (not user replaceable)
Memory size Typical 20 logging sessions of 10 weeks with 10-minute intervals2
Logging period Averaging period Recommended 20 sessions 1 session
1 second 3 hours 2.5 days
5 seconds 15 hours 12 days
10 seconds 28 hours 24 days
30 seconds 3.5 days 10 weeks
1 minute 7 days 20 weeks
5 minutes 5 weeks 2 years
10 minutes 10 weeks > 2 years
15 minutes 3.5 months > 2 years
30 minutes 7 months > 2 years2
Interfaces
USB-A File transfer via USB flash drive, firmware updates 
Max. current: 120 mA
Use of WiFi adaptor requires a powered USB hub and WiFi adaptor with Realtek RTL8188CUS chipset (recommended adaptors are detailed in the operator manual)
USB-mini Data download device to PC
Extension port Accessories
Voltage Inputs
Number of inputs 4 (3 phases and neutral)
Maximum input voltage 1000 Vrms, CF 1.7
Input impedance 10 MΩ
Bandwidth (-3 dB) 2.5 kHz
Scaling 1:1, 10:1, 100:1, 1000:1 and variable
Measurement category 1000 V CAT III / 600 V CAT IV
Current Inputs
Number of inputs 3, mode selected automatically for attached sensor
Input voltage Clamp input: 500 mVrms / 50 mVrms; CF 2.8
Rogowski coil input 150 mVrms / 15 mVrms at 50 Hz, 180 mVrms / 18 mVrms at 60 Hz; CF 4; all at nominal probe range
Range 1 A to 150 A / 10 A to 1500 A with thin iFlex flexible current probe, 12 inches
3 A to 300 A / 30 A to 3000 A with thin iFlex flexible current probe, 24 inches
6 A to 600 A / 60 A to 6000 A with thin iFlex flexible current probe, 36 inches
40 mA to 4 A / 0.4 A to 40 A with 40 A clamp i40s-EL
Bandwidth (-3 dB) 1.5 kHz
Scaling 1:1 and variable
Auxiliary Inputs
Number of inputs 2
Input range 0 to ±10 VDC, 1 reading/s
Scale factor Format: kx + d user configurable
Displayed units User configurable (7 characters, for example, °C, psi, or m/s)
2. The number of possible logging sessions and logging period depends on user requirements.
Environmental Specifications
Operating temperature -10°C to +50°C (14°F to 122°F)
Storage temperature -20°C to +60°C (−4°F to 140°F)
Operating humidity 10°C to 30°C (50°F to 86°F) max. 9% RH
30°C to 40°C (86°F to 104°F) max. 75% RH
40°C to 50°C (104°F to 122°F) max. 45% RH
Operating altitude 2000 m (up to 4000 m derate to 1000 V CAT II / 600 V CAT III / 300 V CAT IV)
Storage altitude 12,000 m
Enclosure IP50 in accordance with EN60529
Vibration MIL 28800E, Type 3, Class III, Style B
Safety IEC 61010-1: Overvoltage CAT IV, Measurement 1000 V CAT III / 600 V CAT IV, Pollution Degree 2
EMI, RFI, EMC EN 61326-1: Industrial
Electromagnetic compatibility Applies to use in Korea only. Class A equipment (industrial broadcasting and communication equipment)
Radio frequency emissions IEC CISPR 11: Group 1, Class A
Temperature coefficient 0.1 x accuracy specification / °C
General Specifications
Color LCD display 4.3-inch active matrix TFT, 480 pixels x 272 pixels, resistive touch panel
Warranty
1730 and power supply Two-years (battery not included)
Accessories One-year
Calibration cycle Two-years
Dimensions
1730 19.8 x 16.7 x 5.5 cm (7.8 x 6.6 x 2.2 in)
Power supply 13.0 x 13.0 x 4.5 cm (5.1 x 5.1 x 1.8 in)
1730 with power supply attached 19.8 x 16.7 x 9 cm (7.8 x 6.6 x 3.5 in)
Weight 1730 1.1 kg (2.5 lb)
Power supply 400 g (0.9 lb)
External protection Holster, Kensington lock slot
1500-12 iFlex Flexible Current Probe Specifications
Measuring range 1 to 150 A AC / 10 to 1500 A AC
Nondestructive current 100 kA (50/60 Hz)
Intrinsic error at reference condition3 ±0.7% of reading
Accuracy 1730 + iFlex ±(1% of reading + 0.02% of range)
Temperature coefficient over operating temperature range 0.05% of reading / °C 0.09% of reading / °F
Working voltage 1000 V CAT III, 600 V CAT IV
Probe cable length 305 mm (12 in)
Probe cable diameter 7.5 mm (0.3 in)
Minimum bending radius 38 mm (1.5 in)
Output cable length 2 m (6.6 ft)
Weight 115 g
Transducer cable material TPR
Coupling material POM + ABS/PC
Output cable TPR/PVC
Operating temperature -20°C to +70°C (-4°F to 158°F) temperature of conductor under test shall not exceed 80°C (176°F)
Temperature, non-operating -40°C to +80°C (-40°F to 176°F)
Relative humidity, operating 15% to 85% non-condensing
IP rating IEC 60529:IP50
Warranty One-year

Permintaan Penawaran